5E-MAG6700 Proximate Analyzer (Thermogravimetric Analyzer, TGA) is used to determine the moisture, ash, volatile matter in coal, coke, biomass, limestone and iron ore, food and feedstuffs by thermogravimetric method. It can also determine residual ash and slag combustibles in accordance with ASTM, which is widely applied in power plants, coal mines, metallurgy, food industry, chemical industry, commercial inspection, scientific research, education, etc. It could work both simultaneously and independently.

Please leave your E-mail address before download the product PDF. Your cooperation is appreciated.

Warm prompt: Please make sure you entered a correct E-mail address. We will send you relevant information and product quotation regularly.

  • Features
  • Specification
  • Spare parts
  • Case
Standard Configuration:Computer  
Printer  
Main analyzer: Part I and Part II  
Air compressor
Volatile matter crucibles  
Ash Crucible 
Weighing rod base
 

Weighing rod  

Three-core power socket and plug/20A 
Crucible lid for volatile matter 
Glass spoon(small size)  
Standard Reference Material(GBW)
Tool kit

Features:High Stability and Reliability1. Dual furnace which could work both simultaneously and independently. Unique isolation board to ensure accurate result of volatile matter. 
2. Patented technology of heat-resistant to ensure stability and reliability. 3. Built-in balance ensures unbeatable accuracy, comply with thermogravimetric method.
3. Built-in balance ensures unbeatable accuracy, comply with thermogravimetric method.
4. Unique design to ensure the stability of heating efficiency. 
5. Using blank crucible to calibrate the influence of the thermal buoyancy. 

Convenient Operation
1.The operator is limited to just adding sample to a crucible.
2. Reference method and customized method available. 
3. Capable to load sample continuously and singly.
4. Real time display of balance reading, which makes it convenient to control the precision of sample weight.

Cost Saving
Available for analysis without oxygen and nitrogen by configured air pump, except for lignite.

Good Safety
Automatically heating off when open the ash furnace cover.

Model 5E-MAG6700 Proximate Analyzer - TGA
Conforms to Method ASTM D3173/D3174/D3175/DD7582, ISO 17246, GB/T 30732, GB/T 212, ISO 18123/18122/18134
Max. Sample Loading 19
Furnace Dual furnace
Analysis Time ≤120mins for 19 samples
Sample Mass 0.8-1.2g recommended / up to 5g
Temp. Range Up to 1050
Temp. Control Precision ± 2
Precision of Balance 0.1mg
Power Supply Single phase, AC220V±10%, 50/60Hz, Part I: ≤ 4 kW
Part II: ≤ 5 kW
Net Weight Part I: 80kg Part II: 50kg
Dimensions (L×W×H) Part I: 550mm×580mm×890mm Part II: 550mm×580mm×530mm