Sample Preparation
Crush & Grind the sample to size ≤ 1mm, then insert the sample into furnace under 815℃ to burn completely ,then grind the ash to the size ≤ 0.1mm
Instrument
 
Instrument Settings
 
Set the temperature of M/VM/A in software before analysis;
Set the keeping time of M/VM/A in software before analysis;
Set the temperature rising speed of VM/A in software before analysis;
Special Modifications to Unit
None
Procedure
1. Weigh 0.6g±0.1g for Moisture and Ash in M&A part according to the instruction manual. Make sure sample in crucible is flattened before you start analysis.
2. Weigh 0.6g±0.1g for Moisture and Ash in VM part according to the instruction manual. After sample weighing, cover the lid on crucible.
3. click start , results will be given automatic after analysis.
Typical Results
    
        
            | Sample Name | CEN/ISO | MAG6700 | Difference between MAG6700 and oven & Muffle furnace | 
        
            | Drying oven | Muffle Furnace | 
        
            | Mad(%) | Ad(%) | Mad(%) | Ad(%) | Mad(%) | Ad(%) | 
        
            | 31 | 8.14 | 4.02 | 7.94 | 4.18 | -0.2 | 0.16 | 
        
            | 51 | 10.23 | 2.04 | 10.16 | 2.23 | -0.07 | 0.19 | 
        
            | 54 | 11.17 | 14.95 | 11.24 | 15.66 | 0.07 | 0.71 | 
        
            | 61 | 9.14 | 16.29 | 9.18 | 16.73 | 0.04 | 0.44 | 
    
        
    
        
            | Sample Name | MAG6700 | CEN | Difference between MAG6700 and CEN | 
        
            | Vd% | Vd% | 
        
            | 61 | 66.16 | 66.66 | -0.5 | 
        
            | 62 | 83.11 | 83.11 | 0 | 
        
            | 75 | 80.16 | 80.07 | 0.09 | 
        
            | 88 | 83.1 | 83.31 | -0.21 |