Contact us now for your demandinfo@ckic.net
Carbon Hydrogen Nitrogen Determination in Coal and Coke
Sample Preparation
Crush & Grind the sample to size ≤ 0.2(GB) / 0.212(ISO) / 0.25mm (ASTM). Air dried Moisture should be tested by oven or TGA in case of calculating dry basis value
Instrument
5E-CHN2200 Elemental Analyzer
Instrument Settings
Oxygen Flow (SV4): 3L/min;
Oxygen Flow (SV5): 0.8L/min;
Oxygen Flow (SV6): 0.3L/min;
Main Furnace Temperature: 950 ℃;
Secondly Furnace Temperature: 850 ℃
Nominal Sample Weight: 80 mg;
Analysis Time: 6 minutes per analysis;
Special Modifications to Unit
None
Method Approval
GB/T 30733 ASTM D5373 AS 1038.6.4 ISO 29541
Procedure
1. Prepare instrument as outlined in the operator’s instruction manual (perform maintenance and leak checks, etc.);
2. Purging and analyze blanks until instrument is stable (nitrogen results will reach a plateau);
3. Calibrate by CRM EDTA;
4. (4-5) blanks and (1-2) waste samples to condition the instrument
5. 3 samples of calibration substance to check or carry out the calibration;
6. Run 9 sample analyses
7. 2 samples of calibration substance to check the calibration;
8. Repetition of point 5 to point 6 until all samples are processed;
Note
Do Calibrating refer to operator’s instruction manual for details.
Code | Name | Mass(g) | Cd(%) | Hd(%) | Nd(%) |
1 | Coal 1 | 0.0801 | 82.98 | 4.27 | 1.80 |
2 | Coal 1 | 0.0816 | 82.93 | 4.20 | 1.76 |
3 | Coke 3 | 0.0792 | 90.43 | 4.03 | 0.95 |
4 | Coke 3 | 0.0813 | 90.54 | 3.96 | 1.01 |